Corrected QT interval (QTc)

All four corrections in common use, because they disagree and the rate determines which is trustworthy.

Also called QTc, Bazett, Fridericia

Values

Result

Enter the values on the left. The answer appears here as you type, and nothing is sent anywhere.

Check the result against an independent reference before acting on it. A calculator agreeing with itself is not a check.

The formula

  • Bazett: QTc = QT ÷ √RR
  • Fridericia: QTc = QT ÷ RR^(1/3)
  • Framingham: QTc = QT + 154 × (1 − RR)
  • Hodges: QTc = QT + 1.75 × (heart rate − 60)
  • RR interval (seconds) = 60 ÷ heart rate

Where it comes from

Bazett HC, Heart 1920;7:353-70. Fridericia LS, Acta Med Scand 1920;53:469-86. Sagie A et al. (Framingham), Am J Cardiol 1992;70(7):797-801. Hodges M et al., J Electrocardiol 1983;16:17-24.

ALL FOUR are shown. Bazett is the one most monitors print and the one that performs worst away from 60 bpm — it over-corrects when fast and under-corrects when slow. Showing only Bazett would be showing the least reliable of the four without saying so.

What it does not account for

The arithmetic is the easy part. These are the reasons a correct number can still be the wrong number to act on.

  • At heart rates away from 60, Bazett and the others diverge substantially. At 100 bpm Bazett can read 20–30 ms longer than Fridericia, which is the difference between acting and not.
  • None of them is reliable in atrial fibrillation, where there is no single RR interval. Averaging several beats is the usual approach and is a clinical judgement, not arithmetic.
  • Bundle branch block and ventricular pacing widen the QRS and so the QT. Correction formulas for that exist and are not implemented here.
  • Measuring the QT is the hard part and no formula helps with it. Automated measurements are frequently wrong, particularly with low-amplitude T waves or a U wave.
  • No threshold is printed. The commonly quoted limits differ by sex, change with drugs and electrolytes, and are a clinical reading rather than an output.

No result is saved

There is no “save to chart” here, deliberately. A derived value stored beside its inputs can disagree with them — and when it does, nothing about the stored number says which is wrong or when its inputs were measured. SendClinic calculates values like these from the measurements in the chart each time they are shown, rather than keeping a copy that can go stale.